Choice of lamps for your QUV or Q-Sun aging speakers
Labomat Essor guides you here in the choice of lamps for your UV aging enclosures QUV models or Xenon models Q-Sun
AMETEK BROOKFIELD
ASCOTT ANALYTICAL EQUIPMENT LTD
ATP ENGINEERING BV
BINDER GmbH
BUCKLEYS (UVRAL) LTD
COATMASTER
DAKOTA ULTRASONICS
DEFELSKO CORPORATION
FORMAT MESSTECHNIK GmbH
HACH LANGE
HANGON
HANNA INSTRUMENTS FRANCE
HILDEBRAND
KERN & SOHN GmbH
LABOMAT ESSOR
LAUDA
LENETA COMPANY
MITUTOYO FRANCE
Q-LAB CORPORATION
RHOPOINT INSTRUMENTS
RK PRINT COAT INSTRUMENTS Ltd
SIGMUND LINDNER GmbH
TABER INDUSTRIES
TAYLOR HOBSON LIMITED
TESTO
TQC BV
VEOLIA WATER STI
VERIVIDE
WALLACE
X-RITE EUROPE GMBH
What is the difference between QFS-40 and UVB-313EL lamps?
QFS-40 and UVB-313EL lamps are both used in QUV accelerated weathering testers to provide fluorescent UV light that includes a significant UVB component for faster testing. The diagram below shows that both have similarly-shaped spectral power distributions. How do these lamps differ, and which one is best for use in your QUV?

The QFS-40 lamp was developed to meet the specifications of SAE J2020, which, when written, had a hardware-specific requirement to use this lamp. That hardware-specific requirement has been removed in SAE J2020 and in most standards that are based on SAE J2020.
UVB-313EL lamps were developed later, and operate with the QUV’s SOLAR EYE irradiance control system to provide a more consistent and repeatable level of irradiance than using QFS-40 lamps in a QUV/basic, due to lamp aging, maintenance, and environmental conditions. UVB-313EL lamps are significantly less expensive than QFS40 lamps because far greater volumes of UBV-313EL are manufactured.
If your QUV has SOLAR EYE irradiance control, you should use UVB-313EL lamps and set the irradiance to 0.48 W/m2 at 310 nm for SAE J2020 and standards based on it, even if those standards have not been updated to include the UVB-313 lamp. This is the most cost-effective and best-performing option when irradiance control is available.
If your QUV does not have irradiance control (QUV/basic or old style models), you should continue to use QFS-40 lamps to maintain comparable irradiance levels to what is intended by a test standard or historical practice. UVB‑313EL lamps used in QUV testers without irradiance control will generate irradiance that is significantly higher than QFS-40 lamps in the same tester.
If your QUV does not have irradiance control (QUV / base or older style models), you should continue to use QFS-40 lamps to maintain irradiance levels comparable to what is expected by a test standard or historical practice. UVB-313EL lamps used in QUV testers without irradiance control generate radiation that is significantly higher than QFS-40 lamps in the same tester.
Labomat Essor guides you here in the choice of lamps for your UV aging enclosures QUV models or Xenon models Q-Sun
Should I Be Concerned with Color Change or with Polymer Degradation ?
Why should I do weathering or light stability testing at all ? What is my ROI ?
I've never done weathering testing, but I'm interested. What's next ?
What is the Best Way to Accelerate Moisture Uptake in a Weathering Test ?